25. Press ENTER touch pad to confirm setting. Display shows:
26. REPEAT WASHING screen allows supervisor to choose if
WASHING phase (WASH segment followed by RINSE segment)
is to be performed once or twice during SINGLE endoscope
processing cycle WITH WASHING or DOUBLE endoscope
processing cycle WITH WASHING (see Table 5-1, Cycle
Description Chart).
Press SCROLL arrows (up or down) to enable or disable
REPEAT WASHING feature (YES = Operator can elect whether
or not to REPEAT WASHING phase; NO = Operator cannot elect
whether or not to REPEAT WASHING phase).
27. Press ENTER touch pad to confirm setting. Display shows:
28. WATER SAMPLING screen allows opening of chamber door
just before draining sub-phase of RINSE 2 phase (refer to
S
4.10, W
ECTION
ATER
Description Chart, for more details).
Press SCROLL arrows (up or down) to enable (YES) or disable
(NO) WATER SAMPLING option.
NOTE: When cycle is completed, water sampling option is
automatically cleared by processor. To perform another Water
Sampling, supervisor must set water sampling option to YES
again.
29. Press ENTER touch pad to confirm setting. Display shows:
CHEMICAL WARNING
30. CHEMICAL WARNING option is set to YES to warn operator
when remaining chemical volume is low. Only approximately five
more cycles can be processed before Bag-in-a-Box is empty
and
Fault
08
or
, for details on fault causes and
T
ROUBLESHOOTING
Cycle may be run without any notification or alarm. Purpose of
chemical warning is to advise operator that Bag-in-a-Box
container needs to be changed soon.
Press SCROLL arrows (up or down) to enable (YES) or disable
(NO) CHEMICAL WARNING feature.
Operator Manual
REPEAT WASHING
YES
__ blinking
WATER SAMPLING
YES
__ blinking
, and Table 5-1, Cycle
S
T
AMPLING
EST
YES
__ blinking
09
is
triggered
Cycle and Control Value Programming
(see
S
7,
ECTION
solutions).

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Reliance EPS Endoscope Processing System and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

This manual is also suitable for:

P920011-649

Table of Contents