5.2.2 Changing Cycle Values

5-4
920011-649
then:
COPYRIGHT (C) 2011
then:
MM-DD-YY
3. Press CHANGE VALUES touch pad to enter Supervisor mode.
Display shows for example:
SCROLL TO MODIFY
and printer records:
===========================
===========================
=
===========================
Changes affect only the selected cycle. Repeat procedure to
change values of other cycles.
Change Cycle Values displays are used for following functions (see
Table 5-1, Cycle Description Chart, for available value ranges):
DISPLAY
Recirculation Time
Print Cycle
1. Access Supervisor mode as explained in S
A
S
CCESSING
UPERVISOR
SCROLL TO MODIFY
2. Press SCROLL arrows (up or down) to choose SINGLE or
DOUBLE.
NOTE: It is also possible to select cycle by pressing directly on
corresponding cycle touch pad.
Operator Manual
STERIS CORP
READY
HH:MM XM
SINGLE
__ blinking
HH:MM:SS
XM MM-DD-YY
SUPERVISOR MODE
FUNCTION
To set recirculation time for WASHING
segments WASH 1 and 2 (from 5 to
10 minutes) and AIR PURGE (from 4 to
30 minutes).
To print values of currently modified cycle.
. Display shows:
M
ODE
SINGLE
__ blinking
Cycle and Control Value Programming
=
5.2.1,
ECTION

Advertisement