FAULT
Alarm sounds, and display shows:
FAULT 37
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM
---------------------------------------
=
FAULT 37
Alarm sounds, and display shows:
FAULT 38
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM
---------------------------------------
=
FAULT 38
Alarm sounds, and display shows:
FAULT 39
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM
---------------------------------------
=
FAULT 39
*Service charges may be incurred. Consult your warranty for details.
Troubleshooting
MM-DD-YY
=
MM-DD-YY
=
MM-DD-YY
=
Operator Manual
POSSIBLE CAUSE AND CORRECTION
PROTECTION RTD DEFECTIVE
Try to resume cycle by pressing START touch pad. If
situation reoccurs, call STERIS.*
FRESH DRY CHEMISTRY CONTAINER NOT DETECTED
A new Reliance dry chemistry container was not inserted in
Reliance CDS prior to cycle start. Cycle is aborted.
Start new cycle after ensuring a new Reliance dry
chemistry container is installed in Reliance CDS and
ensuring latch is closed on Reliance CDS. If situation
reoccurs, call STERIS.*
DRY
CHEMISTRY
DEFECTIVE
Try to resume cycle by pressing START touch pad. If
situation reoccurs, call STERIS.*
CONTAINER
DETECTION
7-17
920011-649