Cycle and Control Value Programming
31. Press ENTER touch pad to confirm setting. Display shows:
32. AUTOMATIC D-SHORT screen allows operator to activate or
deactivate second D-SHORT starting after 36 hours of inactivity.
Press SCROLL arrows (up or down) to enable (YES) or disable
(NO) AUTOMATIC D-SHORT feature.
33. Press ENTER touch pad to confirm selection. Display shows:
Press SCROLL arrows (up or down) to toggle between YES (if
optional Automated Leak Test Assembly is present; automated
leak test at cycle start) or NO (no automated leak test).
NOTE: Leak test is automatically set to YES through a jumper in
the connection.
34. Press ENTER touch pad to confirm selection. Display shows:
Press SCROLL arrows (up or down) to toggle between KLENZYME
or PROLYSTICA.
35. Press ENTER touch pad to confirm selection. Display shows:
36. DUPLICATE PRINTOUT feature is set to YES to allow operator to
generate a duplicate of last cycle printout. After each cycle in
progress, display asks if a duplicate printout is needed.
Press SCROLL arrows (up or down) to enable (YES) or disable
(NO) DUPLICATE PRINTOUT feature.
37. Press ENTER touch pad to confirm selection. Display shows:
Operator Manual
AUTO D-SHORT
YES
__ blinking
LEAK TEST
YES
__ blinking
ENZYME TYPE
KLENZYME
__ blinking
DUPLICATE PRINT?
YES
__ blinking
MULTIPLE PRINTS?
YES
__ blinking
5-13
920011-649
Need help?
Do you have a question about the Reliance EPS Endoscope Processing System and is the answer not in the manual?
Questions and answers