4.8.6 Completing Cycle

4-38
920011-649
1. After four minutes of air purge (adjustable from 4 to 30 minutes in
Supervisor mode), buzzer sounds and display shows:
It is now possible to unload processor or let air purge phase
continue. If door is not opened once set air purge time has
elapsed, air purge continues for up to a total of 60 minutes.
2. Press STOP touch pad to stop air purge and unload processor.
Display shows:
and printer records:
===========================
STOP PRESSED
SAFETY DOOR LOCK
DISENGAGED
===========================
Processor cools for 30 seconds. Then display shows:
Processor drains for 20 seconds.
3. After draining is completed, buzzer sounds and display shows:
4. Printer records complete cycle printout (see Figures 4-24 and
4-25 to view an example of a typical printout).
If EXTENDED PRINTOUT feature is enabled in Miscellaneous
menu of Supervisor mode, there is additional space on
printout to allow operator to write down additional information
(refer to S
ECTION
setting procedure and see Figures 4-24 and 4-25 to view an
example of an extended printout).
If MULTIPLE PRINTOUTS feature is enabled in Miscellaneous
menu of Supervisor mode, printer automatically generates
number of printouts requested in Supervisor mode for each
endoscope (refer to S
V
, for setting procedure).
ALUES
NOTE: If a DOUBLE endoscope processing cycle is running, the
number of printouts doubles.
Operator Manual
CYCLE COMPLETE
PRESS STOP
COOLING
PLEASE WAIT
HH:MM:SS XM
HH:MM:SS XM
DRAINING
PLEASE WAIT
CYCLE COMPLETE
READY TO UNLOAD
5.2.3, S
M
ETTING
ISCELLANEOUS
5.2.3, S
ECTION
ETTING
V
, for
ALUES
M
ISCELLANEOUS
Processing Instructions