FAULT
Alarm sounds, and display shows:
FAULT 43
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM
---------------------------------------
=
FAULT 43
FAULT 44 – 50 Hz Processors.
FAULT 45 – 50 Hz Processors.
Alarm sounds, and display shows:
FAULT 46
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM
---------------------------------------
=
FAULT 46
*Service charges may be incurred. Consult your warranty for details.
Troubleshooting
MM-DD-YY
=
MM-DD-YY
=
Operator Manual
POSSIBLE CAUSE AND CORRECTION
SINGLE/DOUBLE CYCLE OVER TEMPERATURE
Try to resume cycle by pressing START touch pad. If
situation reoccurs, call STERIS.*
Refer to Operator Manual - Information Supplement For
50 Hz Processors (P920501-322) for specific information.
Refer to Operator Manual - Information Supplement For
50 Hz Processors (P920501-322) for specific information.
LEAK TEST PORT 1: OVERPRESSURE
Press ALARM REPLY touch pad to silence audible alarm.
Pressure inside endoscope 1 over maximum 4.33 psig
(0.30 bar).
Cycle aborted; users are required to restart a processing
cycle. If unsuccessful, call STERIS.*
Possible cause:
Leak test port 1 pressure transducer not calibrated.
7-19