Chapter 6 Specifications - Thermo Scientific Sorvall MTX 150 Instruction Manual

Table of Contents

Advertisement

Specifications
Model
Maximum speed
Maximum RCF*
Speed control accuracy
Rotor temperature
control/display accuracy
Set speed
Vacuum system
Noise level
Maximum heat
dissipation into room
Cooling method
Screen display and operation
Dimensions
Weight
Power requirement***
Ambient temperature
* RCF is an acronym for relative centrifugal force.
** Please note that the LCD panel may contain a few dead or stuck pixels.
*** The voltage to be used is the one that you specified when purchasing the micro-ultracentrifuge.
Thermo Scientific
Sorvall MTX 150 micro-ultracentrifuge
150,000 rpm
1,048,680 xg
(S140-AT)
±50 rpm (5,000 rpm to maximum speed)
±2°C set temperature is from 0°C to 40°C
5,000 rpm to maximum speed in increments of 1,000 rpm
Oil rotary vacuum pump and oil diffusion pump combined
Ultimate vacuum: below 0.6 Pa (0.005 Torr)
45 dB (A scale)
(measured 1m in front of the instrument)
0.7 kW or less
Thermo-module cooling (CFC-free)
**Color touch-sensitive LCD
Width: 590 mm; Depth: 582 mm; Height: 408 mm
97 kg
Single phase: 110 or 120 Vac+/-10%; 50/60 Hz; 15 A
208, 220, 230, or 240 Vac+/-10%; 50/60 Hz; 8 A
Ambient temperature for performance: 10 °C to 30 °C
Ambient temperature for operation: 5 °C to 35 °C
6
Sorvall MTX 150
6-1

Advertisement