Thermo Scientific Sorvall MTX 150 Instruction Manual page 60

Table of Contents

Advertisement

2
Process
Touch panel operation
3
When RCF
or RCF
max
column,
SPEED
SPEED
touch the
show SPEED in the SPEED column. Press the
SPEED column and enter the SPEED value you
want to convert into the RCF value with the
keypad display.
(e.g.) Rotor: S150AT
SPEED: 100,000 rpm
Press
1
0
button to enter the setting.
4
Press the
SPEED
SPEED
turn the SPEED column to the RCF
column or the RCF
2. How to set an RCF value
Select the desired rotor and enter an RCF value, and the micro-ultracentrifuge will calculate, set, and
display the speed. Shown below is the procedure
2-42
Sorvall MTX 150
is displayed in the
avg
button to
RCF
RCF
0
, and press the Enter
button to
RCF
RCF
max
column.
avg
Screen display and notice
• See
"Setting run conditions"
RCF
column: The maximum RCF value for the rotor set
max
speed is displayed.
RCF
column: The average RCF value for the rotor set
avg
speed is displayed.
on
page 2-5
SPEED column
• The RCF value of the
rotor displayed in the
ROTOR indicator is
displayed on the setting
value display field of the
RCF column.
This is the RCF value for
the SPEED value entered in
step 3.