Summary of Contents for Thermo Scientific MaxQ 2000
Page 1
MaxQ 2000 Open-Air Platform Shaker Operating Manual 057-278-00 (7004310) Rev. 13 Visit us online to register your warranty www.thermoscientific.com/labwarranty...
Page 2
Added note to hyperterminal config information - pg 4-4 28316 3/6/12 Added platform mntg plate replacement instructions to Maintenance section 11/1/11 Added missing line to Step 12 on page 4-4 _ lost during transfer 27615 7/22/11 Revised warranty Thermo Scientific MaxQ 2000 Shaker...
Page 4
Always use the proper protective equipment (clothing, gloves, goggles, etc.) 4 Always dissipate extreme cold or heat and wear protective clothing. 4 Always follow good hygiene practices. 4 Each individual is responsible for his or her own safety. Thermo Scientific MaxQ 2000 Shaker...
Page 5
We can also provide you with a quotation on our Extended Warranty for your Thermo Scientific products. Whatever Thermo Scientific products you need or use, we will be happy to discuss your applications. If you are experiencing technical problems, working together, we will help you locate the problem and, chances are, correct it yourself...over the telephone without a service...
Safety Information Section 1 Your Thermo Scientific MaxQ 2000 Shaker has been designed with function, reliability, and safety in mind. It is your responsibility to install it in conformance with local electrical codes. For safe operation, pay attention to the alert signals (Warnings, Cautions, Notes) throughout the manual.
1hr. Mutable Alarms None None without deactivating the actual without deactivating the actual alarm condition by depressing alarm condition by depressing any key any key * Interface cables not to exceed 9.8’ (3m) in length Thermo Scientific MaxQ 2000 Shaker...
Page 11
1hr. be silenced for a period of 1hr. Mutable Alarms without deactivating the actual without deactivating the actual alarm condition by depressing alarm condition by depressing any key any key MaxQ 2000 Shaker Thermo Scientific...
Page 12
Optional Platform Dimensions in. (cm) Catalog Number L x W 30100 Universal 13” x 11” (33 x 27.9 cm) 30110 Universal 18” x 18” (45.7 x 45.7 cm) 30106 Universal 18” x 24” (45.7 x 60.9 cm) Thermo Scientific MaxQ 2000 Shaker...
Section 2 General Specifications The Thermo Scientific MaxQ 2000 Series bench top, orbital, open-air shakers are available in one of two control configurations: • SHKA2000 and SHKA2000-1CE incorporate a mechanical timer for the control of time and speed, adjustable with rotary dials. Analog tachometer displays speed in RPM, verifying accuracy of speed setting.
1. Power Switch: Turns power on and off to shaker 2. Speed Control: Sets platform rotation speed 3. Speed Tachometer: Analog display of platform rotation speed 4. Time(r): Allows user to choose either continuous or timed operation Thermo Scientific MaxQ 2000 Shaker...
16. Set Time: Illuminates when time is being set 17. Timer/Elapsed: Allows the user to choose elapsed time operation, Elapsed, or timed operation, Timer 18. Hours/Minutes Membrane Switch: Allows user to choose timing operation in either hours or minutes MaxQ 2000 Shaker Thermo Scientific...
Place shaker near an electrical outlet that matches the unit nameplate requirements. Allow clearance around the unit for free air convection, accessory attachments and user convenience. Thermo Scientific MaxQ 2000 Shaker...
4 thumbscrew knobs to secure platform to the mounting plate. Caution Do not operate shaker with an unbalanced load. Platforms should be loaded for optimum stability and operation. Do not lift shaker by the platform. MaxQ 2000 Shaker Thermo Scientific...
Installation mounting holes on the platform. There are two cutouts on each side of the rack. 2. Secure the rack to the platform with mounting screws provided with the rack. Thermo Scientific MaxQ 2000 Shaker...
Warning Do not use shaker to mix flammable materials or where the transfer of mechanical energy to glass could cause glass breakage. Note Shaker will not operate if the timer is in the Off position. Thermo Scientific MaxQ 2000 Shaker...
6. Use the UP or DOWN arrow keys to set the shaker speed to match the tachometer’s readout. 7. Press STOP key to enter the new speed value. 8. Press START key to exit the calibration mode. MaxQ 2000 Shaker Thermo Scientific...
Microsoft Windows 98 or newer operating system. RS232 Interface Port Note RS232 Interface Port is for output only. Interface cables must not exceed 9.8ʼ (3m) in length. Thermo Scientific MaxQ 2000 Shaker...
10. Turn shaker off and connect computer (COM 1) to shaker (COM PORT) with DB-9 serial printer cable. 11. Start HyperTerminal by clicking on “Max Q Shaker.” 12. Power up shaker. Shaker will display speed, time and temperature at one minute intervals. MaxQ 2000 Shaker Thermo Scientific...
(pressed in). 3. While it is slowly orbiting, slightly tighten each of the A screws in Figure 1. Repeat this procedure by slightly tightening all the B screws, then complete procedure by tightening the C screws. Thermo Scientific MaxQ 2000 Shaker...
Thermo. Warning Disconnect plug from electrical outlet before attempting any maintenance or repair of this unit. Note The shaking mechanism is equipped with sealed ball bearings which do not require further lubrication or adjustment. MaxQ 2000 Shaker Thermo Scientific...
Check for power to motor Replace motor if defective Platform doesn't rotate Replace if worn, broken or Check drive belt or has erratic speed slipped off pulley A-Class, check for power to speed control Replace if defective Thermo Scientific MaxQ 2000 Shaker...
If the dealer cannot process your request, then contact our Technical Services Department. Prior to returning any materials, contact our Technical Services Department for a “Return Materials Authorization” number (RMA). Material returned without an RMA number will be refused. MaxQ 2000 Shaker Thermo Scientific...
Need help?