Operation
How to use the option function
Process
Touch panel operation
1
Check that the desired rotor is displayed in
the ROTOR indicator of the Run screen. Go
to step 3 when the desired rotor is
displayed. When the desired rotor is not
displayed, press the ROTOR column to show
the ROTOR CATALOG screen.
2
Press the column of the rotor you want to
see the RCF value. The column turns from
white to blue. Check that the pressed
column turns blue and press the Enter
button.
3
When "SPEED" is displayed in the
column, press the
SPEED
SPEED
button to show
RCF
RCF
"RCF
" ("RCF
max
column. Press the RCF
and enter the SPEED value you want to convert
into the SPEED value with the keypad display.
(e.g.) Rotor: S150AT
RCF
: 850x1,000 xg
max
Press
0
button to enter the setting.
Thermo Scientific
") in the RCF
(RCF
)
avg
max
avg
(RCF
) column
max
avg
8
5
, and press the Enter
Screen display and notice
• See
"Setting run conditions"
ROTOR indicator
The pressed column
turns from white to
blue.
on
page
2-5.
RCF
column
max
Sorvall MTX 150
2
2-43