Thermo Scientific Sorvall MTX 150 Instruction Manual page 79

Table of Contents

Advertisement

2
Figure 2-15. Backlight Setting screen
5. Vacuum standby setting
The conditions of the waiting for the vacuum of the rotor acceleration state can be
specified.
1. Normal:
When the vacuum level returns from medium to low during acceleration at 5,000 rpm or higher speed, the
micro-ultracentrifuge keeps on running at that speed until the vacuum level becomes medium again.
2. No Waiting Acceleration:
The micro-ultracentrifuge keeps on accelerating when it is accelerating at 5, 000 rpm or higher speed
even after the vacuum level returns from medium to low.
Press either the Normal or No Waiting Acceleration button and make sure that the selected button is
surrounded with a green frame. Then press the x button or the CUSTOM screen tab to store the
setting.
Figure 2-16. Vacuum Standby Setting screen
Thermo Scientific
Sorvall MTX 150
2-61