1
further (dual CPU overspeed preventive mechanism).
The first CPU detects overspeed and performs control and display. Should the rotor be set to a
speed exceeding the permissible speed, this CPU detects an alarm massage of "SPEED" from the
low speed range (about 2,000 rpm), and stops the rotor.
(But this second CPU does not display an alarm message, because it is not connected to the
display-performing CPU. Should the alarm device be activated, pressing START button will not
run the instrument. Turn off the POWER switch, wait for several minutes, turn the POWER
switch on again, then pressing START button.)