Process
Touch panel operation
4
Press the
SPEED
SPEED
the RCF
(RCF
max
column.
Note 1. To shift from the SPEED column to the RCF column, press the
2. This micro-ultracentrifuge is not capable of checking rotors. Select the correct rotor.
Especially when you wish to determine a speed based on an RCF value, selecting the wrong rotor
and thus setting an RCF value exceeding the capacity of your rotor will cause the system to calculate
and set a speed exceeding the maximum allowable speed. (However, this micro-ultracentrifuge has
an overspeed detector, which prevents an overspeed run.)
3. When you run this micro-ultracentrifuge at a speed determined based on an RCF value, there
may occur a slight error (of up to 2%) between the set RCF value and the actual value, because the
speed setting is set in increments of 1,000 rpm.

User login

After logging in, individual users can control the operation history of the micro-ultracentrifuge.
1. Procedure for user login
Process
Touch panel operation
1
Press the USER column on the Run screen.
2-44
Sorvall MTX 150
button to turn
RCF
RCF
) column to the SPEED
avg
Screen display and notice
Screen display and notice
• The SPEED value of
the rotor displayed in the
ROTOR indicator is
displayed on the setting
value display field of the
SPEED column.
This is the SPEED value
for the RCF value entered
in step 3.
button.
SPEED
SPEED
RCF
RCF
USER column