Thermo Scientific Sorvall MTX 150 Instruction Manual page 86

Table of Contents

Advertisement

2
3. Actual Run Timer setting
Press either the Enable or Disable button and make sure that the selected button is surrounded with a
red frame. Then press the x button or the ADMIN screen tab to store the setting.
Figure 2-20. Actual Run Timer Setting screen
4. Date and time
Figure 2-21. Setting at time screen
2-68
Sorvall MTX 150
You can select whether to give displays in English or Japanese by pressing this
button.
<Actual Run Timer is enabled.>
When you press the Enable button, the timer starts counting after reaching the
set speed.
<Actual Run Timer is not enabled.>
When you press the Disable button, the timer starts counting immediately after
the start of operation.
Use this function to precisely set the current time setting of the internal clock.
Set an exact date and time for RTC operation.