Thermo Scientific Sorvall MTX 150 Instruction Manual page 30

Table of Contents

Advertisement

2
Process
Touch panel operation
8
Press the VACUUM button.
9
Take out the rotor.
Note When the rotor chamber is vacuumed insufficiently before starting operation or the ambient
temperature is low (10°C or below), the vacuum waiting time at 5,000rpm may become longer.
Also, during acceleration up to the set speed, the instrument may become a vacuum waiting state.
Therefore, before starting operation, vacuum the rotor chamber sufficiently (approx. 15 minutes)
and then press the VACUUM button and check that the displayed vacuum is at a high vacuum
level (3 indicator).
2-12
Sorvall MTX 150
Screen display and notice
• The vacuum stops, the air leak valve becomes activated, and
the rotor chamber reaches atmospheric pressure.
• The door unlocks, and is able to be opened and closed.
• Stop the rotor completely before taking it out.