Figure 5-1. Clearance for micro-ultracentrifuge
2. Electrical power requirement
An emergency switch (circuit breaker) should be installed that turns off the main power supply in the
event of malfunctioning. (The desirable installation location of the emergency switch is outside the
room or near the exit.)
Do not position an object so that it is difficult to disconnect the power cord from the outlet. If you do
so, you cannot disconnect the power cord from the outlet when you observe some abnormality.
Your micro-ultracentrifuge can operate on one of the following six power voltages:
110 or 120 Vac (50/60 Hz, 15 A)
208, 220, 230, or 240 Vac (50/60 Hz, 8 A)
The voltage requirement for your micro-ultracentrifuge is mentioned on the rectangular marking plate
(rating plate) affixed near the power cord connector, which is visible in the rear panel of the
micro-ultracentrifuge. Be sure to read the marking plate before plugging the micro-ultracentrifuge.
If the voltage requirement does not match the voltage of the available power source, quit plugging and
call your Thermo Fisher Scientific representative.
3. Leveling
(1) Turn the leveling screws with a spanner.
5-2
Sorvall MTX 150
8 cm or more
13 cm
or more
CAUTION Your micro-ultracentrifuge can be damaged if connected to a wrong voltage.
Check the voltage before plugging the micro-ultracentrifuge into a power source.
WARNING Your micro-ultracentrifuge must be grounded properly.
13 cm
or more