I-Drive Power Activation/Deactivation - LINET Sprint 200 Instructions For Use And Technical Description

Emergency stretcher without scales / with i-drive power and without i-drive power
Hide thumbs Also See for Sprint 200:
Table of Contents

Advertisement

13.11.4 i-Drive Power Activation/Deactivation

To prepare the i-Drive Power wheel for use:
Press the mains switch located on the undercarriage cover to the ON position.
Press green Drive pedal to the lower position.
The braked i-Drive Power Wheel will lower.
To activate the i-Drive Power:
Press i-Drive Power Wheel Activation Button
flashing. Place your hand on the Safety Sense touch sensor to use the i-Drive Power.
To retract the i-Drive Power Wheel:
Press or lift a pedal to leave all Brake pedals and Drive pedals unpressed in their neutral middle position or brake the
Sprint 200.
To deactivate the i-Drive Power:
It is recommended to fold the Foldable infusion stand with i-Drive Power Control Panel down.
Press the mains switch located on the undercarriage cover to the OFF position.
Fig. Position of the i-Drive Power Mains Switch
Fig. i-Drive Power Mains Switch with Label
74
located on the Main Control Panel. The green ON LED
will be
D9U001ES2-0101_02