Set Clock

5-12 | Thermox CG1000 Oxygen Analyzer
Set the current time and date and define a recovery duration.
During the recovery duration the sensor switches from measuring
calibration gas readings to measuring process readings without affect-
ing the analog output or triggering of alarms.
NOTE

Set Time and Date

Set the date or time for the analyzer's internal clock.
1.
From the Calibrate key menu, select Set Clock.
2.
From the Set Clock menu option, select Set Time & Date.
3.
Enter the new time, then press Enter. As you type the new time, it
overwrites the Hr: Mn text. The CG1000 uses military time so 16:00
equals 4:00 PM., and 04:00 equals 4:00 AM. If the current time is cor-
rect, press Enter without changing the value.
4.
As you type the new date, it overwrites the mo/da/yr text on the
display. Enter the month, the day, the year, then press Enter. If the
current date is correct, press Enter without changing the value.

Recovery Duration

Define a recovery time so the analyzer can return to reading process gases
after reading calibration gases without affecting the analog outputs or the
triggering of alarms. This menu option applies to both calibrations and
verifications.
1.
From the Calibrate key menu, select Set Clock.
2.
From the Set Clock menu option, select Recovery Duration.
Enter the recovery time, then press Enter. If you want to use the cur-
3.
rently displayed recovery time, press Enter without entering any text.
The time is: YY:XX
Enter new time: HR: Mn
The date is: YY/XX/ZZ
Enter date: Mo/Da/Yr
Cal Recovery = xx:yy
Enter new time: Mn:Sc