Interfaces - Dräger Julian Operating Instructions Manual

Anesthesia workstation
Hide thumbs Also See for Julian:
Table of Contents

Advertisement

Technical Data
Measuring Functions

Interfaces

SpO
measurement, optional
2
(light absorption)
Display range
Accuracy (adults)
between 70 and 100 % SpO
between 50 and 70 % SpO
between 0 and 50 % SpO
Accuracy (neonates)
between 70 and 95 % SpO
between 0 and 70 % SpO
between 95 and 100 % SpO
Trigger point
Pulse rate
Accuracy
Sensors
Type
Wavelengths
Acoustic pulse signal
Interfaces
Analog output
Plug connector
PAW
Flow
CO
2
Ground
COM1 record (printer)
Connector
Pin assignment
COM2 (MEDIBUS 1)
Connector
Pin assignment
COM3 (MEDIBUS 2)
Connector
Pin assignment
142
0 to 100 % SpO
Better than ±2 % SpO
2
Better than ±3 % SpO
2
Not specified
2
Better than ±3 % SpO
2
Not specified
2
Not specified
2
Once per heartbeat
20 to 250 per minute
±2 per minute
Compatible with Nellcor sensors
Oxisensor, Oxiband and Durasensor
660 nm (red)
920 nm (infrared)
A tone is produced for each detected pulse, tone pitch proportional
to oxygen saturation.
9-pin sub-D, electrical isolation 1.5 kV
0 to 100 cmH
0 to 150 L/min; 0 to 5 V=; PIN 6
0 to 10 kPa; 0 to 5 V=; PIN 9
PIN 1, 4, 7
9-pin sub-D, electrical isolation 1.5 kV
1 shield
2 TxD
3 RxD
5 GND
9-pin sub-D, electrical isolation 1.5 kV
1 shield
2 TxD
3 RxD
5 GND
9-pin sub-D, electrical isolation 1.5 kV
1 shield
2 TxD
3 RxD
5 GND
2
2
2
2
O; 0 to 5 V =; PIN 3
2