Operation
Selecting Ventilation Modes
Alarms effective in CMV mode:
Alarm
Limits
In CMV mode
SpO
ON
2
_
ON
ON
ª
Pulse
_
ON
EtCO
ON
2
_
ON
FiCO
ON
2
>
MV
ON
_
ON
FiO
ON
2
_
ON
Fi Hal.
ON
_
ON
Fi Iso.
ON
_
ON
Fi Enf.
ON
_
ON
Fi Des.
ON
_
ON
Fi Sev.
ON
_
ON
PAW
ON
_
ON
Apnea
pressure
ON
Apnea
flow
ON
Apnea
CO
ON
2
– – : The alarm limit is deactivated as factory default.
To set the values of the alarm limits during operation see
page 72.
To set the default alarm limits see page 98.
52
Factory default
setting
– –
92
120
50
50 mmHg
– –
5 mmHg
– –
3.0
– –
20
1.5
– –
2.3
– –
3.4
– –
12.0
– –
3.4
– –
40
8
Active after
15 seconds
Active after
15 seconds
Active after
30 seconds