Dräger Julian Operating Instructions Manual page 41

Anesthesia workstation
Hide thumbs Also See for Julian:
Table of Contents

Advertisement

O
flush
2
1 Press »O
+« button.
2
2 The breathing bag inflates.
3 O
flow is audible.
2
Safety O
2
1 Turn Safety O
control valve counterclockwise.
2
2 The breathing bag fills up.
1 Set Safety O
back to » 0 «.
2
Preparing Julian for the self test
1 Close the Y-piece = plug firmly onto the cone.
2 Ensure that the sample line is not connected to the
patient circuit.
3 If necessary, seal the sample-line connector with
a luer-lock cap.
4 Set APL valve to MAN and adjust to 30 cmH
Ensure that Safety O
flow control is set to » 0 «.
2
3
O.
2
Preparation
Checks of Readiness for Operation
1
2
1
2
2
4
3
1