Dräger Julian Operating Instructions Manual page 55

Anesthesia workstation
Hide thumbs Also See for Julian:
Table of Contents

Advertisement

Alarms effective in PCV mode:
Alarm
Limits
In PCV mode
SpO
2
_
ª
Pulse
_
EtCO
2
_
FiCO
2
>
MV
_
FiO
2
_
Fi Hal.
_
Fi Iso.
_
Fi Enf.
_
Fi Des.
_
Fi Sev.
_
PAW
_
Apnea
pressure
Apnea
flow
Apnea
CO
2
– – : The alarm limit is deactivated as factory default.
To set the values of the alarm limits during operation see
page 72.
To set the default alarm limits see page 98.
Factory default
settings
ON
– –
ON
92
ON
120
ON
50
ON
5 mmHg
ON
– –
ON
5 mmHg
ON
– –
ON
3.0
ON
– –
ON
20
ON
1.5
ON
– –
ON
2.3
ON
– –
ON
3.4
ON
– –
ON
12.0
ON
– –
ON
3.4
ON
– –
ON
40
ON
8
ON
Active after
15 seconds
ON
Active after
15 seconds
ON
Active after
30 seconds
Operation
Selecting Ventilation Modes
55