Setting Default Alarm Limits - Dräger Julian Operating Instructions Manual

Anesthesia workstation
Hide thumbs Also See for Julian:
Table of Contents

Advertisement

Configuring in Standby Mode
Setting Default Values
Interface
Choice between configuration of COM2 or COM3.
Baud rate
Transmission rate (variable, see Instructions for Use
of the equipment to be connected).
Parity
This display is fixed for MEDIBUS and is shown for
information purposes only.
Data bits
This display is fixed for MEDIBUS and is shown for
information purposes only.
Stop bits
This display is fixed for MEDIBUS and is shown for
information purposes only.

Setting Default Alarm Limits

Turn the rotary knob to select the line »alarm limits«
and press to confirm.
The »alarm limits« menu appears in the left-hand
field.
Display (example):
Select required parameter by turning the rotary knob
and press to confirm.
The "alarm limits" menu appears.
In the menu, select the required alarm limit by turning
the rotary knob and press to confirm. The default
alarm limit is highlighted against a white background.
Set and confirm the value with the rotary knob. The
value is then displayed in normal mode.
The cursor bar jumps to the next default alarm limit.
Set the next default alarm limit as described above.
The following alarm limits can be configured:
SpO
/pulse
Pulse rate
SpO
2
2
CO
/O
etCO
FiCO
2
2
2
Flow/pressure MV
PAW
Anesth. agent Fi Iso. Fi Hal.
MAN/SPONT
etCO
FiCO
2
alarms
_
>
The etCO
, FiCO
, FiO
2
2
can be configured »ON« or »OFF« in standby mode for
the transition from MAN/SPONT. When the alarm limits
are set to »ON«, the value is transferred from automatic
ventilation mode.
96
FiO
2
2
Fi Sev. Fi Enf. Fi Des.
FiO
MV
2
2
>
_
and MV
alarms
2
Standby / Configuration
MAC-config.
calibration
SpO
/pulse
2
100
CO
/O
2
2
SpO
90
2
Flow/pressure
120
50
Anaesthetic agent
MAN/SPONT alarms
default settings
pulse tone
0 1 2 3 4
5
6 7 8 9
alarm sound
1 2 3
4
5 6 7 8 9
scale
adult
child
parameters
record
interfaces
alarm limits
curves
basic configurations
IPPV default settings
PCV default settings
fresh gas delivery