Monitoring
Configuring During Operation
Record
This function is used to determine the type of event
that triggers an entry in the numerical trend or an
online printout on the connected logging printer.
Select the line »record« with dial knob, press to
confirm.
The »record« menu appears on the left-hand side.
Display (example):
Time interval-started (min)
A list entry is activated after a fixed time interval in
minutes.
NiBP-started
An entry is made after every NiBP measurement with
new measured values. These values are read from an
external monitor which is connected to Julian via
MEDIBUS interface.
If no Dräger NiBP monitor is connected, no NiBP
entries will be activated.
warning-started
An entry is made whenever a Warning message is
triggered.
caution-started
An entry is made whenever a Caution message is
triggered.
Turn dial knob to select event triggering of entries,
press to confirm.
Activating monitor defaults
This function is used to activate the default monitoring
settings configured in standby mode which are active
whenever the system is switched on.
Alarm limits are not affected and remain at their
current settings.
Select the line »activate monitor defaults« with dial
knob, press to confirm, see page 92, "Setting Default
Values".
80
CMV
MAC-config.
alarms
calibration
PAW
list entry
40
time
interval (min)
1
2
5
10
NiBP started
on
off
warn. started
on
off
caution started
on
off
20
printer entry
time
interval (min)
1
2
5
10
NiBP started
on
off
0
warn. started
on
off
on
off
caution started
Configuration menu for List and Printer Entries
Fresh Gas O + N O
2
2
P
V
max
T
O %
L/min
cmH O
mL
2
2
50
1.00
25
600 10
settings
0 1 2 3 4
5
6 7 8 9
pls. tone
alarm sound
1 2 3
4
5 6 7 8 9
off
scale
adult
child
parameter
record
activate monitor defaults
off
Freq.
T
:T
T :T
PEEP
IP I
I
E
bpm
%
cmH O
2
1:2.0
10
0