Foreword; Product Description; Design; Function - Otto Bock Kenevo 3C60 Instructions For Use Manual

Hide thumbs Also See for Kenevo 3C60:
Table of Contents

Advertisement

Available languages

Available languages

1 Foreword

INFORMATION
Date of last update: 2021-01-13
Please read this document carefully before using the product and observe the safety
notices.
Obtain instruction from the qualified personnel in the safe use of the product.
Please contact the qualified personnel if you have questions about the product or in case of
problems.
Report each serious incident related to the product to the manufacturer and to the relevant
authority in your country. This is particularly important when there is a decline in the health
state.
Please keep this document for your records.
The product "Kenevo 3C60/3C60=ST" is referred to as the product/prosthesis/knee joint below.
These instructions for use provide you with important information on the use, adaptation and
handling of the product.
Only put the product into use in accordance with the information contained in the accompanying
documents supplied.

2 Product description

2.1 Design

The product consists of the following components:

2.2 Function

This product features a microprocessor-controlled switch between the stance phase and swing
phase and a microprocessor-controlled stance phase.
The microprocessor uses the measurements of an integrated sensor system as a basis to control
a hydraulic unit that influences the damping behaviour of the product.
These sensor data are updated and evaluated 100 times per second. As a result, the behaviour of
the product is adapted to the current motion situation (gait phase) dynamically and in real time.
Thanks to the microprocessor-controlled stance phase, the knee joint can be individually adapted
to your needs.
The product can be individually adapted to your needs with adjustment software.
Through the adjustment software, it is possible to choose from three activity modes that make the
various functions of the product available. This permits optimum adaptation of the product to the
52
1. Connection of the knee joint to a trans­
femoral socket or other prosthesis compon­
ent
2. Optional flexion stops
3. Battery and cover caps
4. Hydraulic unit
5. Receiver of the inductive charging unit