Directives And Manufacturer's Declaration; Electromagnetic Environment - Otto Bock 17KO1 Series Instructions For Use Manual

C-brace joint unit
Hide thumbs Also See for 17KO1 Series:
Table of Contents

Advertisement

Available languages

Available languages

Appendices

17.3 Directives and manufacturer's declaration

17.3.1 Electromagnetic environment

This product is designed for operation in the following electromagnetic environments:
Operation in a professional healthcare facility (e.g. hospital, etc.)
Operation in areas of home healthcare (e.g. use at home, use outdoors)
Observe the safety notices in the section "Information on proximity to certain areas" (see Page 71).
Electromagnetic emissions
Interference measure­
ments
HF emissions according to
CISPR 11
Harmonics according to
IEC 61000-3-2
Voltage fluctuations/flicker
according
IEC 61000-3-3
Electromagnetic interference immunity
Phenomenon
Electrostatic discharge
High-frequency
electro­
magnetic fields
Magnetic fields with rated
power frequencies
Electrical
fast
transi­
ents/bursts
Surges
Line against line
Conducted
interference
induced by high-frequency
fields
Voltage drops
Voltage interruptions
112
Compliance
Group 1/class B
Not applicable – power
below 75 W
Product meets the require­
to
ments of the standard.
EMC basic standard or
test procedure
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-8
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-6
IEC 61000-4-11
IEC 61000-4-11
Electromagnetic environment directive
The product uses HF energy exclusively for its internal
functioning. Its HF emissions are therefore very low,
and interference with neighbouring electronic devices is
unlikely.
Interference immunity test level
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air,
10 V/m
80 MHz to 2.7 GHz
80% AM at 1 kHz
30 A/m
50 Hz or 60 Hz
± 2 kV
100 kHz repetition rate
± 0.5 kV, ± 1 kV
3 V
0.15 MHz to 80 MHz
6 V in ISM and amateur frequency bands between
0.15 MHz and 80 MHz
80% AM at 1 kHz
0% U
; 1/2 period
T
At 0, 45, 90, 135, 180, 225, 270 and 315 degrees
0% U
; 1 period
T
and
70% U
; 25/30 periods
T
Single phase: at 0 degrees
0% U
; 250/300 periods
T
C-Brace joint unit 17KO1=*

Advertisement

Chapters

Table of Contents
loading

Table of Contents