17.3 Directives and manufacturer's declaration

17.3.1 Electromagnetic environment

This product is designed for operation in the following electromagnetic environments:
Operation in a professional healthcare facility (e.g. hospital, etc.)
Operation in areas of home healthcare (e.g. use at home, use outdoors)
The customer or user of the product must ensure that it is operated in such an environment.
Observe the safety notices in the section "Information on proximity to certain areas" (see Page 14).
The following tables outline the test levels of the tests to be carried out. The higher value applies if there is a differ­
ence between the operating environments listed in the tables below.
Table 4 - casing
Phenomenon
Electrostatic discharge
High
frequency
electro­
magnetic fields
High
frequency
electro­
magnetic
fields
in
immediate vicinity of wire­
less
communication
devices
Magnetic fields with rated
power frequencies
Table 5 - alternate current gate for the supply input
Phenomenon
Electrical
fast
transi­
ents/bursts
Surges
Line against line
Surges
Line against ground
Conducted
interference
induced by high frequency
fields
C-Leg 3C98-3/3C88-3
EMC basic standard or
test procedure
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-3
the
IEC 61000-4-8
EMC basic standard or
test procedure
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-5
IEC 61000-4-6
Interference immunity test level
Professional healthcare
facility
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air,
3 V/m
80 MHz to 2.7 GHz
80 % AM at 1 kHz
See Table 9
30 A/m
50 Hz or 60 Hz
Interference immunity test level
Professional healthcare
facility
± 2 kV
100 kHz repetition rate
± 0.5 kV, ± 1 kV
± 0.5 kV, ± 1 kV, ± 2 kV
3 V/
0.15 MHz to 80 MHz
6 V
in
ISM
frequency
bands between 0.15 MHz
and 80 MHz
80 % AM at 1 kHz
Appendices
Environments in areas of
home healthcare
10  V/m
80 MHz to 2.7 GHz
80 % AM at 1 kHz
Environments in areas of
home healthcare
3 V/
0.15 MHz to 80 MHz
6 V in ISM and amateur
frequency bands between
0.15 MHz and 80 MHz
80 % AM at 1 kHz
45