Battery charger disconnected
Beep sig­
Vibration
nal
signal
1 x short
1 x short
3 x short
Battery charge level
Charger
Battery is charging.
The on time of the LED indicates the current battery charge level.
The on time of the LED gets longer as the charge level increases. It only flashes briefly at the start
of the charging process and stays on continuously at the end of the charging process.
Battery is fully charged, or the temperature fell above/below the permissible range during charging.
Check current charge level (see Page 19).

14.3 Directives and manufacturer's declaration

14.3.1 Electromagnetic environment

This product is designed for operation in the following electromagnetic environments:
Operation in a professional healthcare facility (e.g. hospital, etc.)
Operation in areas of home healthcare (e.g. use at home, use outdoors)
The customer or user of the product must ensure that it is operated in such an environment.
Observe the safety notices in the section "Information on proximity to certain areas" (see Page 14).
The following tables outline the test levels of the tests to be carried out. The higher value applies if there is a differ­
ence between the operating environments listed in the tables below.
Table 1 – Electromagnetic emissions for all devices and systems
Interference measurements
HF emissions according to CIS­
PR 11
HF emissions according to CIS­
PR 11
Harmonics
according
IEC 61000-3-2
Voltage fluctuations/flicker accord­
ing to IEC 61000-3-3
Table 4 - casing
Phenomenon
Electrostatic discharge
High
frequency
electro­
magnetic fields
Kenevo 3C60/3C60=ST
Event
Self-test completed successfully. Product is operational.
Maintenance note
Conduct the self-test again by connecting/disconnecting the battery charger. If the
beep signal sounds again, product maintenance should be carried out by an author­
ised Ottobock Service Centre.
The product can be used without restrictions. However, vibration signals may not be
generated.
Compliance
Group 1
Class B
to
N/A
N/A
EMC basic standard or
test procedure
IEC 61000-4-2
IEC 61000-4-3
Electromagnetic environment dir­
ective
The product uses HF energy exclus­
ively for its internal functioning. Its
HF emissions are therefore very low,
and interference with neighbouring
electronic devices is unlikely.
The product is intended for use in
other than residential facilities and in
facilities connected directly to a pub­
lic low-voltage network that also sup­
plies residential buildings.
Interference immunity test level
Professional healthcare
facility
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air,
3 V/m
80 MHz to 2.7 GHz
80 % AM at 1 kHz
Appendices
Environments in areas of
home healthcare
10  V/m
80 MHz to 2.7 GHz
80 % AM at 1 kHz