Directives And Manufacturer's Declaration; Electromagnetic Environment - Otto Bock Genium X3 Instructions For Use Manual

Bionic prosthetic system
Hide thumbs Also See for Genium X3:
Table of Contents

Advertisement

17.3 Directives and manufacturer's declaration

17.3.1 Electromagnetic environment

This product is designed for operation in the following electromagnetic environments:
Operation in a professional healthcare facility (e.g. hospital, etc.)
Operation in areas of home healthcare (e.g. use at home, use outdoors)
Observe the safety notices in the section "Information on proximity to certain areas" (see page 21).
Table 1 – Electromagnetic emissions for all devices and systems
Directives and manufacturer's declaration – electromagnetic interference
The product 'Genium X3 - Bionic Prosthetic System 3B5-X3 / Genium X3 3B5-2' is designed for operation in an
electromagnetic environment as specified below. The customer or user of the product 'Genium X3 - Bionic Pros­
thetic System 3B5-X3 / Genium X3 3B5-2' must ensure that the device is operated in such an environment.
Interference measurements
HF emissions according to CIS­
PR 11
HF emissions according to CIS­
PR 11
Harmonics
according
IEC 61000-3-2
Voltage fluctuations/flicker accord­
ing to IEC 61000-3-3
Table 2 – Electromagnetic interference immunity for all devices and systems
Directives and manufacturer's declaration – electromagnetic interference immunity
The product 'Genium X3 - Bionic Prosthetic System 3B5-X3 / Genium X3 3B5-2' is designed for operation in an
electromagnetic environment as specified below. The customer or user of the product 'Genium X3 - Bionic Pros­
thetic System 3B5-X3 / Genium X3 3B5-2' must ensure that the device is operated in such an environment.
Interference immunity
test
Electrostatic
discharge
(ESD)
according
IEC 61000-4-2
Electrical
fast
transi­
ent/bursts
according
IEC 61000-4-4
Surges
according
IEC 61000-4-5
Genium X3 - Bionic Prosthetic System 3B5-X3/3B5-X3=ST,Ge­
nium X3 3B5-2/3B5-2=ST
Compliance
Group 1
Class B
to
N/A
N/A
IEC 60601 test level
± 6 kV discharge on con­
to
tact
± 8 kV discharge through
air
± 2 kV for power lines
to
±1 kV for input and output
lines
to
±1 kV
phase-to-phase
voltage
± 2 kV even voltage
Electromagnetic environment dir­
ective
The product 'Genium X3 - Bionic
Prosthetic System 3B5-X3 / Genium
X3 3B5-2' uses HF energy exclus­
ively for its internal functioning.
Therefore its HF emissions are very
low and interference with neighbour­
ing electronic devices is unlikely.
The product 'Genium X3 - Bionic
Prosthetic System 3B5-X3 / Genium
X3 3B5-2' is intended for use in oth­
er than residential facilities and in
facilities connected directly to a pub­
lic low-voltage network that also sup­
plies residential buildings.
Compliance level
± 6 kV discharge on con­
tact
± 8 kV discharge through
air
N/A
N/A
Appendices
Electromagnetic environ­
ment directive
Floors should be wood,
concrete or ceramic tile. If
the floor covering is made
of synthetic material, the
relative humidity must be at
least 30 %.
The quality of the supply
voltage should be equival­
ent to a typical commercial
or hospital supply.
The quality of the supply
voltage should be equival­
ent to a typical commercial
or hospital supply.