Linear Mode Procedure

1. Set the power level on the synthesized sweeper to –30 dBm.
2. Set the analyzer by pressing the following keys:
3. Adjust the amplitude on the synthesized sweeper until the marker
Under these analyzer conditions, the sweep time is 1.7 seconds.
NOTE
Therefore, the marker amplitude updates are fairly slow when
adjusting the synthesizer output power.
Chapter 3
Synthesized
Sweeper Amplitude
(dBm)
Reference + _____
Reference + (10 dB)
Reference + (20 dB)
Reference + (–10 dB)
Reference + (–20 dB)
Reference + (–30 dB)
Reference + (–40 dB)
,
Sweep
Sweep (Cont)
,
AMPLITUDE
–30 dBm
,
AMPLITUDE
+21.76 dBmV
Scale Type (Lin)
,
AMPLITUDE
More 1 of 2
,
AMPLITUDE
More 1 of 2
,
Marker
Off
amplitude on the analyzer reads –30 dBm +/– 0.10 dB. Enter the
synthesized sweeper power level as the Synthesized Sweeper
Amplitude reference in
Minimum
Analyzer
Marker ∆
(dB)
Amplitude
(dB)
0 (Reference)
0 (Reference)
8.60
18.60
–11.40
–21.40
–31.40
–42.0
(50 Ω Input)
(75 Ω Input)
,
,
Amptd Units
dBm
,
,
Amptd Units
dBmV
Table 3-5 on page
3-36.
Functional Testing
Reference Level Accuracy
Maximum
(dB)
0 (Reference)
11.40
21.40
–8.60
–18.60
–28.60
–38.0
(50 Ω Input)
(75 Ω Input)
3-35