Procedure (3.01 Ghz To 6.0 Ghz) E4404B, E4405B, E4407B, E4408B - Agilent Technologies ESA-E Series User Manual

Esa spectrum analyzers, 9 khz-1.5/3.0/6.7/13.2/26.5 ghz
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Functional Testing
Displayed Average Noise Level
Procedure (3.01 GHz to 6.0 GHz) E4404B, E4405B,
E4407B, E4408B
1. Press
,
System
Power On/Preset
analyzer. Wait for the preset routine to finish. Set the analyzer by
pressing the following keys:
,
FREQUENCY
Start Freq
,
Stop Freq
6.0 GHz
,
AMPLITUDE
–70 dBm
,
Attenuation (Man)
,
BW/Avg
1 MHz
Video BW Auto Man
2. Press the following keys on the analyzer:
Single
,
BW/Avg
Average (On)
Single
Wait until AVG 3 is displayed to the left of the graticule (the
analyzer will take 3 sweeps and then stop).
3. Press
Peak Search
column (g) in
Table 3-1
4. Press the following keys on the analyzer:
,
Sweep
Sweep (Cont)
BW/Avg, Average (Off)
,
BW/Avg
Resolution BW (Auto)
Video BW (Auto)
,
SPAN
50 kHz
FREQUENCY
5. Press
. Set the center frequency of the analyzer to the
Center Freq
frequency recorded in the Measured Frequency column (g) in
Table 3-1
for 3.01 GHz to 6.0 GHz.
6. Press the following keys on the analyzer:
,
BW/Avg
1 kHz
Video BW Auto Man
Single
Wait for the sweep to finish.
3-14
,
Preset (Factory)
,
3.01 GHz
0 dB
,
10 kHz
,
,
3
Enter
. Record the frequency in the Measured Frequency
for 3.01 GHz to 6.0 GHz.
,
30 Hz
,
on the
Preset
Chapter 3

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents