7. Press the following keys on the analyzer:
,
BW/Avg
1 kHz
Video BW Auto Man
Single
Wait for the sweep to finish.
8. Press the following keys on the analyzer:
,
Display
Display Line (On)
Adjust the display so that it is centered on the average trace noise,
ignoring any residual responses.
9. Record the display line amplitude setting as TR Entry 1 (TR Entry
16 for a 75 Ω Input) in
less than the Maximum for the appropriate input impedance.

Procedure (501 MHz to 1.0 GHz) E4401B and E4411B

1. Press the following keys on the analyzer:
,
Sweep
Sweep (Cont)
,
FREQUENCY
Start Freq
,
Stop Freq
1.0 GHz
,
BW/Avg
1 MHz
Video BW Auto Man
2. Press the following keys on the analyzer:
Single
,
BW/Avg
Average (On)
Single
Wait until AVG 3 is displayed to the left of the graticule (the
analyzer will take 3 sweeps, then stop).
3. Press
Peak Search
Frequency column (b) in
Chapter 3
,
30 Hz
Table
3-1. The Average Noise Level should be
,
501 MHz
,
10 kHz
,
,
3
Enter
, and record the frequency in the Measured
Table 3-1
for 501 MHz to 1.0 GHz.
Functional Testing
Displayed Average Noise Level
3-7