Frequency Readout Accuracy

Test Limits
Frequency Readout Accuracy is equivalent to the following equation:
±
(
×
×
0.25% span
+
5% RBW
See
Table 3-1 on page 20
The frequency readout accuracy is measured in several spans and center frequencies that allow both
internal analyzer synthesizer modes and prefilter bandwidths to be tested. Frequency reference error is
eliminated by using the same frequency standard for the analyzer and signal source.
Item
Adapter, Type-N(m),
Frequency: 10 MHz to 1.51 GHz
to 3.5 mm(f)
VSWR: < 1.1:1
Adapter, 3.5 mm(f) to
Frequency: 10 MHz to 1.51 GHz
3.5 mm(f)
VSWR: < 1.1:1
Cable, 3.5 mm, 1
Frequency: 10 MHz to 1.51 GHz
meter
VSWR: < 1.4:1
Cable, BNC, 120 cm
Frequency: 10 MHz
Synthesized Sweeper
Frequency: 10 MHz to 1.51 GHz
Figure 3-1.
Frequency Readout Accuracy Test Setup
×
+
2 Hz
+
0.5
horizontal resolution
for actual values.
Critical Specification
(for this test)
)
Recommended
Agilent Model
1250-1744
83059B
11500E
10503A
PSG
19