C.3

Setpoint offset operating mode (Diff.set/actual)

It is possible to apply an offset value to the temperature, which is provided by an external temperature sensor and to
process it as the set value.
The bath temperature can therefore be operated, for example, -15 °C below the temperature of a reactor measured by
the external temperature sensor.
You activate the soft-key bar by pressing any key.
off
extern Pt100
extern USB
 OK
ESC
Offset value
Min: -500.0
 OK
ESC
By pressing
92 / 124
STOP
Max: 500.0
-15.0
+ / -
(ESC) you are returned to the menu level without any change.
Access to the main menu level is obtained by pressing the enter
key .
Selection and confirmation of 
Setpoint
offset. 
The adjacent menu window appears.
Select and confirm offset source with
The setpoint offset is deactivated with "off".
The entry window appears on selecting the menu point
Diff.set/actual.
The minimum and maximum possible offset values and the current offset
value are displayed.
Change the value with
Single figures can be selected by pressing or .
By pressing
(+/-) the arithmetic sign can be changed.
Confirm your choice with the enter key
ECO Silver
Setup
Control
Offset
source.
or
or .
.
and
.