CAUTIONS
An error occurs if the system is started up immediately after shutdown.
To restart the system including the control cabinet, make sure that the power status LED of the
control cabinet is off, and then press the power switch for the image processing unit.
CAUTIONS
● Do not connect/disconnect the fl at panel sensor to/from the power supply unit or to/from the
power box while the message "Calibrating..." is displayed in the connected devices status after
the system startup. Otherwise, the system does not start up normally, resulting in an error.
the system startup. Otherwise, the system does not start up normally, resulting in an error.
may be displayed in the connected devices status while information on radio wave
strength is being acquired from the fl at panel sensor.
When the Power supply unit and Power box are used, refer to the respective operation manuals.

3.2.2 Shutting Down the System

(When only the fl at panel sensor is used)
1
Confi rm that the equipment is not running. Touch the
processing unit display, and then the
button in the displayed confi rmation window.
The image processing unit will shut down in a few minutes. If the control cabinet is included in the
system, the control cabinet will also turn off automatically.
button at the upper right of the image
button from the displayed menu. Touch the
(1)
(2)
(3)
DR-ID1270 Operation Manual
897N120821E
3
3-9