Appendix A Specifications; Specifications; Reduced Equivalent (Dr-Id1270; Power Supply Conditions - FujiFilm FDR ES DR-ID1270 Operation Manual

Table of Contents

Advertisement

Appendix A
A.1 Specifications
Specifications of the DR-ID1270 are shown below.
A.1.1 Reduced Equivalent (DR-ID1270)
Peak reduced equivalent on the front panel of the flat panel sensor: 0.5 mmAl
A.1.2 Power Supply Conditions
■ Flat panel sensor
Rated voltage : 6-12V
Input current : 2.73A
■ Power supply unit/Power box
Rated voltage : 100-240V~
Input current : 2-0.84A
Frequency
A.1.3 Environmental Conditions
■ Flat panel sensor/DR-ID1270MP/DR-ID1270PB
(1) Operating Conditions
Temperature
Humidity
Atmospheric pressure : 700hPa - 1060hPa
(2) Non-operating Conditions
(Environmental conditions under which power can be supplied)
Temperature
Humidity
Atmospheric pressure : 700hPa - 1060hPa
(3) Storage Conditions
Temperature
Humidity
Atmospheric pressure : 700hPa - 1060hPa
CAUTIONS
● When the flat panel sensor is used in high temperature condition for long period of time, it
may cause image artifacts and/or failure of the device.
● When using the DR-ID 1271SE, DR-ID 1273SE or DR-ID 1275SE, if the temperature is 37ºC
and the humidity is 90% RH (no dew condensation), continuous use of 30 minutes or less is
possible.
Using Manual Mode (energy saving mode) from the image processing unit when the
temperature and humidity are the same allows up to 1 hour of continuous use.

Specifications

: 50-60Hz
: 15°C (15%RH) - 35°C (80%RH)
: 15%RH (15°C) - 80%RH (30°C) (no dew condensation)
: 5°C - 35°C
: 10%RH - 80%RH (no dew condensation)
: -30°C - 50°C
: 10%RH - 90%RH (no dew condensation)
DR-ID1270 Operation Manual