2.4.2 Installation Precautions

■ Using in an open space
Install the fl at panel sensor, the access point and the image processing unit at a distance of less
than 10m (32.8 ft) from each other. If any distance is over 10m (32.8 ft), a wireless communication
error may occur.
■ Using in an X-ray room
The access point should be installed in the X-ray room.
If the access point is not installed properly, wireless communication may become unstable.
If this happens, the messages below may appear.
These messages are only examples.
Example 1
Panel communication error

2.4.3 Precautions for Installing the Access Point (Optional)

Install the optional access point in a place where the operation is not hindered.
When the access point is installed onto a personal computer, be sure that the label attached on the
access point does not face the front side.
In addition, install the access point in an appropriate place to prevent it from colliding with a moving mobile X-ray
unit. If any impact is applied to the optional access point, it may be damaged.
Example 2
Connection to X-ray unit has been disconnected.
DR-ID1270 Operation Manual
2
2-11
897N120821E