Disinfection Instructions
Confirm that the respiratory density of disinfectant including solvent is under legal regulation.
Certain disinfectants may damage health. When using a disinfectant, follow instructions
supplied by the manufacturers.
1
Do not use the following disinfectants or sterilizers at the time of disinfection. Quality,
performance and safety of the equipment cannot be assured.
● Chloric disinfectant which is strongly corrosive to metals and rubber parts.
● Disinfectant whose uses on metals, plastics, and coating are forbidden according to the
instructions supplied with the disinfectant.
● Formalin gas and disinfectant sprays that may get inside the equipment.
Disinfectant ethanol is recommended for disinfection. Carefully read the instructions and
cautions supplied with the disinfectant before use.
For details on the disinfectant, contact a FUJIFILM dealer or the service representatives at the
agency from which you purchased the disinfectant.
For details of the ultraviolet sterilizers, see the each operation manual of DR-ID1271SE/
DR-ID1272SE/DR-ID1273SE/DR-ID1274SE/DR-ID1275SE, Power supply unit or Power box.
If the sensor unit of the flat panel sensor is not disinfected, it may lead secondary infection.
Clean the sensor unit of the flat panel sensor with ethanol for disinfection, etc. for each patient
to prevent infection.
Precautions for Charging the Battery Pack
● Use the battery charger recommended by FUJIFILM Corporation.
For details on operations, refer to the instruction manual for the battery charger.
● Do not charge the battery pack near fire or under strong sunshine. If the built-in protection
mechanisms are activated by a high temperature, the battery pack cannot be charged. Also,
if the built-in protection mechanisms are damaged, the battery pack may be charged with
extremely high current and voltage, and abnormal chemical reactions may occur inside the
battery pack, causing it to overheat, emit smoke, explode or ignite.
● To charge the battery pack, be sure to use the designated battery charger and to observe
the charging conditions specified by FUJIFILM Corporation. If the battery pack is charged in
other conditions (temperature or voltage/current higher than specified, remodeled battery
charger, etc.), the battery pack may be overcharged or charged with extremely high current,
and abnormal chemical reactions may occur inside the battery pack, causing it to overheat,
emit smoke, explode or ignite.
● Immediately stop charging the battery pack, if charging is not completed within the specified
time. Otherwise, the battery pack may overheat, emit smoke, explode or ignite.
● Do not use the flat panel sensor near the power cable.
● Do not use a faulty or broken battery charger or AC adapter.
● Note that the flat panel sensor cannot be charged by using the SE communication cable that
connects the flat panel sensor and the access point (optional).