FujiFilm FDR ES DR-ID1270 Operation Manual page 23

Table of Contents

Advertisement

If a seal that covers a screw peels from the side surface of the
flat panel sensor, contact a FUJIFILM dealer. If the seal is not
attached, artifacts caused by discharge of static electricity may
appear.
To ensure optimal image quality, it is recommended that you do
not use the flat panel sensor near devices (motor, transformer,
1
switching supply, etc.) that generate electromagnetic noise.
To ensure optimal image quality, it is recommended that you do
not place the cables (power cable, communication cable, etc.)
of the equipment near devices (motor, transformer, switching
supply, etc.) that generate electromagnetic noise and their
cables.
Make sure that no liquid enters the flat panel sensor from
around the battery section. In addition, when attaching the
battery pack, make sure that the waterproof packing attached
to the connector terminal of the flat panel sensor is aligned
properly. Otherwise, the flat panel sensor may be damaged.
Do not use a multiple tap connector or extension cable for
powering the devices constituting the system.
Up to five flat panel sensors can be connected. If you intend
to use six or more flat panel sensors, only the first five that
were connected to the image processing unit can be used.
For this reason, when six or more flat panel sensors are
registered, be careful not to use a wrong one, as you may
confuse which flat panel sensor is connected.
Before making an exposure, make sure that the flat panel
sensor identification lamp on the flat panel sensor to be
used and the panel icon selected on the screen of the image
processing unit are the same color.
When a flat panel sensor is communicating with a power
supply unit or access point in a room, if the flat panel sensor
is moved to another room where there is another power
supply unit or access point, communication between the flat
panel sensor and the device in the first room may still be
established. To establish communication between the flat
panel sensor and a device in the second room, connect the
flat panel sensor to the device with the cable. The flat panel
sensor is recognized and wireless communication becomes
available.
Do not place the cable terminal on the floor, as doing so may
cause infection.
Also, clean the cable and the terminal periodically.
Do not insert the flat panel sensor into a CR reader unit.
1-17
DR-ID1270 Operation Manual
897N120821E

Advertisement

Table of Contents
loading

Table of Contents