Warnings for Abnormalities
WARNING
If any of the following occurs, immediately turn off the power of each unit, unplug the power
cable from the outlet, and then contact our official dealer or FUJIFILM Representative.
● When smoke, strange odor, or abnormal sound is present.
● When a foreign object (such as a metal object) or liquid enters the product.
● When the equipment is dropped or hit and is damaged.
Avertissements relatifs aux anomalies
AVERTISSEMENT
Si l'une des conditions répertoriées ci-après se produit, mettez immédiatement chaque unité
hors tension, débranchez le cordon d'alimentation de la prise secteur, puis contactez notre
revendeur agréé ou notre représentant FUJIFILM.
● En cas de présence de fumée, d'une odeur étrange ou d'un bruit anormal.
● En cas de pénétration d'un corps étranger (comme un objet métallique) ou d'un liquide dans
le produit.
● En cas d'endommagement de l'équipement suite à une chute ou à un impact.
Installation Precautions
CAUTIONS
Do not install the equipment in a location with the following conditions.
● Where the temperature changes sharply.
● Close to heat sources such as a heater.
● Where the equipment may be exposed to water due to water leakage or ingress.
● Where corrosive gas may be generated.
● Where there is excessive dust.
● Where the equipment is subject to frequent or excessive vibration/shock.
● Where the equipment is exposed to direct sunlight.
● Where there is no ventilator.
CAUTIONS
When the power supply unit is installed, be sure to use the Retaining bracket for MP or fixing
metal fittings to prevent its overturning.
CAUTIONS
Use the equipment on a flat place. If the equipment falls, it may cause damage to the equipment
or personal injury.
CAUTIONS
When you move the equipment, place it in the cassette storage box of a mobile X-ray unit or hold
it by hand to prevent it from falling. If the cart is used to move the equipment, place it horizontally.
CAUTIONS
For veterinary or mobile applications, contact our official dealer or FUJIFILM Representative.