Maintenance And Inspection - FujiFilm FDR ES DR-ID1270 Operation Manual

Table of Contents

Advertisement

Maintenance and Inspection

1 Maintenance and Inspection Items Assigned to Specified Dealer
For periodical inspection of the equipment and necessary arrangements, consult our official dealer
or local representative.
Periodical Maintenance
Make sure that the periodical maintenance and inspection assigned to our official dealer or
FUJIFILM Representative are performed as specified.
Maintenance and Inspection Items Assigned to Specified Dealer
Periodical Maintenance and Inspection Items
Checking of the image
Checking of the operation record by referring to the error log
Checking of the units
Main Periodical Replacement Parts
Name of Periodical Replacement Parts
Relay (optional)
* It is recommended that the battery pack be replaced, if the battery storage capacity becomes
lower than 60%.
The battery pack should be replaced when the operable time is less than the following.
● DR-ID1271SE/DR-ID1272SE/DR-ID1273SE/DR-ID1274SE/DR-ID1275SE :
100 minutes (1 hour and 40 minutes)
* Refer to the operable time displayed on the image processing unit when the battery pack is fully
charged and no exposure menu is registered.
* Depending on the usage environment, etc., the displayed time is slightly different from the
actual operable time.
120
100
80
60
60
40
20
0
0
30
The cycles of periodical maintenance and inspection and of parts replacement differ depending on
the usage and the daily operation time.
For details, contact our official dealer or FUJIFILM Representative.
(Number of exposures : 90,000)
(Reference) Time for battery pack replacement
Time for replacement:
Less than 100 minutes
(1 hour and 40 minutes)
60
90
120
150
Period
Every year
Every year
Every 2 years
Period
Every 1 years
180
210
240
Operable time [Minute]
DR-ID1270 Operation Manual
897N120821E