FujiFilm FDR ES DR-ID1270 Operation Manual page 18

Table of Contents

Advertisement

Further Information for IEC 60601-1-2 (EN 60601-1-2)
1. Medical electrical equipment needs special precautions regarding EMC and needs to be installed
and put into service according to the EMC information provided in the accompanying documents.
2. Portable and mobile RF communications equipment can affect medical electrical equipment.
3. Information regarding the cable affecting EMC is as follows.
1
Network Cable
* When the distance between the Power box and the DR-ID 1200MC is 2 meters or more, install a
commercially available HUB. In that case, use a cable with a maximum length of 2 meters between the
Power box and HUB, and a cable with a maximum length of 20 meters between HUB and
the DR-ID 1200MC.
4. The use of optional items, transducers and cables other than those specified, with the exception
of transducers and cables sold by FUJIFILM Corporation as replacement parts for internal
components, may result in increased emissions or decreased immunity of the DR-ID1270.
5. The DR-ID1270 should not be used adjacent to or stacked with other equipment.
If adjacent or stacked use is necessary, the DR-ID1270 should be observed to verify normal
operation in the configuration in which it will be used.
6. Basic performance of the equipment and the system
After image data are acquired from the flat panel sensor, data correction is performed by the control
cabinet, and the image is saved in and displayed on the image processing unit.
7. Test items (Tables 1 to 4)
Table 1
The DR-ID1270 is intended for use in the electromagnetic environment specified below.
The customer or the user of the DR-ID1270 should assure that they are used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
1-12
DR-ID1270 Operation Manual
Name
Connected Device
Between the Power supply
unit and the DR-ID 1200MC
Between the Power box and
the DR-ID 1200MC*
Guidance and manufacturer's declaration - electromagnetic emissions
Compliance
Group 1
Class B
Complies
Class B
Complies
897N120821E
Maximum Length
20m (65.6 ft)
2m (6.56 ft)
Electromagnetic environment - guidance
The DR-ID1270 uses RF energy only for their internal function.
Therefore, their RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
The DR-ID1270 is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
General Specification