HEIDENHAIN QUADRA-CHEK ND 1100 Operating Instructions Manual page 90

Table of Contents

Advertisement

11
Setup | Correcting for errors
The MZ Offset field will automatically populate with the difference between the
zero reference location of the standard and the machine zero
Highlight the Station field
The Station must be zero at the beginning of the SLEC procedure.
Enter zero values into the Standard and Observed fields to define the zero
station as the reference
Position the crosshairs over the end of segment 1 on the standard
Highlight the Station field again
Press the Inc soft key to advance the station number to 1
Highlight the Standard field
Enter the Standard value at the end of segment 1
In the example, the value is 10.0000.
Highlight the Observed field
Press the Teach soft key
The system will automatically populate the value measured at the end of the
segment. In the example the Observed value at Station 1 is also 10.0000.
Position the crosshairs over the end of segment 2 on the standard
Highlight the Station field again
Press the Inc soft key to advance the station number to 2
Highlight the Standard field
Enter the Standard value at the end of segment 2
In the example, the value is 20.0000.
Highlight the Observed field
Press the Teach soft key
The system will automatically populate the value measured at the end of the
segment. In the example the Observed value at Station 2 is 19.8000.
Position the crosshairs over the end of segment 3 on the standard
Highlight the Station field again
Press the Inc soft key to advance the station number to 3
Highlight the Standard field
Enter the Standard value at the end of segment 3
In the example, the value is 30.0000.
Highlight the Observed field
Press the Teach soft key
The system will automatically populate the value measured at the end of the
segment. In the example the Observed value at Station 3 is 29.8000.
Position the crosshairs over the end of segment 4 on the standard
Highlight the Station field again
Press the Inc soft key to advance the station number to 4
This is the last station in the example.
Highlight Standard
Enter the Standard value at the end of segment 4
In the example, the value is 40.0000.
Highlight the Observed field
Press the Teach soft key
The system will automatically populate the value measured at the end of the
segment. In the example the Observed value at Station 4 is 40.000.
90
HEIDENHAIN | ND 1100 | Operating Instructions | 05/2018