HEIDENHAIN QUADRA-CHEK ND 1100 Operating Instructions Manual page 86

Table of Contents

Advertisement

11
Example: LEC
In this example of applying LEC, one point at the end of the axis measurement
range is measured using an 80 mm standard.
1
2
3



4


Number
1: Standard length
2: Standard values
3: Observed values
4: Deviation graph
To perform the linear error correction in the LEC menu:
Highlight the LEC menu
Enter the Standard value of the artifact and the Observed value measured by
the product for the measurement axis
Perform LEC on other axes if desired
Press the finish key
The parameters are saved and the Setup menu is displayed
86
The standard and observed values for axes should be 1.000 when no
LEC correction is applied.


Description
Entire 80 mm length is measured
The certified length of the standard
The measured length of the standard
Difference between standard and
observed values (not entered into any
menu)
NOTICE
HEIDENHAIN | ND 1100 | Operating Instructions | 05/2018
Setup | Correcting for errors