13.2 Precautions during SpO
1
Verify the sensor cable fault detection before the beginning of monitoring phase.
Unplug the SpO
message "SpO
2
If the SpO
sensor can not work properly, please reconnect the sensor or change a
2
new one.
3
Do not use the sterile supplied SpO
damaged and return them to the vendor.
4
Prolonged and continuous monitoring may increase the risk of unexpected change of
dermal condition such as abnormal sensitivity, rubescence, vesicle, repressive
putrescence, and so on. It is especially important to check the sensor placement of
neonate and patient of poor perfusion or immature dermogram by light collimation
and proper attaching strictly according to changes of the skin. More frequent
examinations may be required for different patients.
5
Tissue damage may be caused by incorrect application or prolonged measurement
duration using the sensor (more than 4 hours). Inspect the sensor periodically
according to the sensor user manual.
6
The sensor's applicable wavelengths are 660nm of red light and 905 nm of infrared
light.
7
The sensor accords with the ISO 10993-1: 2009 for biocompatibility.
NOTE:
1
Make sure the nail covers the light window. The wire should be on the backside of the
hand.
2
Hand should not be too cold when measuring, and the nail polish should be cleaned
before measuring, or the data accuracy may be affected.
3
SpO
value always displays at the same position. Pulse Rate will display when HR
2
FROM is set to "SpO
4
SpO2 waveform is not proportional to the pulse volume.
5
A functional tester cannot be used to assess SpO

13.3 Monitoring Procedure

1. Switch on the monitor.
2. Attach the sensor to the appropriate site of the patient finger.
3. Plug the connector of the sensor extension cable into the SpO
sensor cable from the socket, and the screen will display the error
2
SENSOR OFF" and the audible alarm is activated.
2
", No ECG signal when HR FROM is set to AUTO.
2
/Pulse Monitoring
2
WARNING
sensors if the packaging or the sensor is
2
accuracy.
2
- 103 -
SpO
socket on the SpO
2
2
Monitoring
2
module.