Patient Monitor User Manual
Applicable
Prompt
Patient Type
Without
VBRADY
pacemaker
Without
VENT
pacemaker
Patient type:
All patients: refers to performing Arr.analysis on patients either with pacemakers or without
pacemakers.
Without pacemaker: refers to performing Arr. Analysis only on the patients without pacemakers.
With pacemaker: refers to performing Arr. Analysis only on the patients with pacemakers.
Prompt message:
Message
ARR LEARNING
NOTE:
Arrhythmia name displays in the Alarm Message Area.

12.9 Measuring RESP

Cardiogenic artifact in impedance respiration monitoring may make it difficult to detect
breaths or may otherwise be counted as breaths. In some instances, the breath rate may
also correspond to the heart rate making it difficult to determine if the signal is due to
breathing or the cardiac cycle. Do not rely on RESP monitoring as the sole method for
detecting cessation of breathing. Follow hospital guidelines and best clinical practices for
apnea detection including monitoring additional parameters that indicate the patient's
oxygenation status, such as etCO
Occurring Condition
VENTRICULAR BRADYCARDIA: The
interval of 5 consecutive ventricular wave
is more than 1000 ms.
VENTRICULAR RHYTHM: The interval
of 5 consecutive ventricular wave ranges
from 600 ms to 1000 ms.
Cause
The QRS template building
required for Arr. Analysis is
in process.
WARNING
and SpO
.
2
2
- 96 -
ECG/ RESP Monitoring
Alarm Level
User-selectable
User-selectable
Alarm Level
No alarm