EDAN iM8 Series User Manual page 164

Patient monitor
Hide thumbs Also See for iM8 Series:
Table of Contents

Advertisement

Patient Monitor User Manual
Range
Accuracy
Resolution
HR averaging method
Method 1
Method 2
Range of Sinus and SV Rhythm
Tachy
Normal
Brady
Range of Ventricular Rhythm
Ventricular Tachycardia
Ventricular Rhythm
Ventricular Bradycardia
Maximum Start-up time for Tachycardia
Ventricular Tachycardia
1 mV 206bpm
Ventricular Tachycardia
2 mV 195bpm
-2.0 mV ~ +2.0 mV
±0.02 mV or 10% (-0.8 mV ~ +0.8 mV), whichever is
greater.
0.01 mV
Normally, heart rate is computed by excluding the
minimum and maximum values from the 12 most recent
RR intervals and averaging the residual 10 RR intervals.
If each of three consecutive RR intervals is greater than
1200ms, then the four most recent RR intervals are
averaged to compute the HR.
ADU: 120 bpm ~ 300 bpm
PED/NEO: 160 bpm ~ 350 bpm
ADU: 41 bpm ~ 119 bpm
PED/NEO: 61 bpm ~159 bpm
ADU: 15 bpm ~ 40 bpm
PED/NEO: 15 bpm ~ 60 bpm
The interval of 5 consecutive ventricular wave is less than
600 ms
The interval of 5 consecutive ventricular wave ranges from
600 ms to 1000 ms
The interval of 5 consecutive ventricular wave is more than
1000 ms
Gain 1.0: 10 s
Gain 0.5: 10 s
Gain 2.0: 10 s
Gain 1.0: 10 s
Gain 0.5: 10 s
Gain 2.0: 10 s
- 155 -
Specifications