A2.3 Electromagnetic Immunity - for EQUIPMENT and
SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture's declaration – electromagnetic immunity
The Patient Monitor is intended for use in the electromagnetic environment specified below.
The customer or the user of Patient Monitor should assure that it is used in such an
environment.
Immunity
IEC/EN 60601 test
test
3 V
Conducted RF
150 kHz to 80 MHz
IEC/EN
61000-4-6
3 V/m
Radiated RF
80 MHz to 2.5 GHz
IEC/EN
61000-4-3
Compliance
level
3 V
rms
3 V/m
Electromagnetic environment -
level
Portable and mobile RF
communications equipment should
be used no closer to any part of the
Patient Monitor, including cables,
than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
rms
3
5 .
=
d
V
1
3
5 .
=
d
E
1
7
=
d
E
1
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in
metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
be less than the compliance level in
each frequency range.
Interference may occur in the
vicinity of equipment marked with
the following symbol:
- 166 -
EMC Information
guidance
P
80 MHz to 800 MHz
P
800 MHz to 2.5 GHz
P