Spo Monitoring Failures; Nibp Monitoring Failures; Temp Monitoring Failures - EDAN iM80 Servise Manual

Patient monitor
Hide thumbs Also See for iM80:
Table of Contents

Advertisement

iM80 Patient Monitor Service Manual
HR
inaccurate,
abnormal, and ST analysis
incorrect.
7.11 SpO
Monitoring Failures
2
Phenomenon
No SpO
waveform.
2
Strong
SpO
waveform
2
interference.
SpO
value is inaccurate.
2

7.12 NIBP Monitoring Failures

Phenomenon
The cuff fails to be inflated.
No NIBP numeric value.
Big error in NIBP numeric
value.

7.13 TEMP Monitoring Failures

Phenomenon
TEMP
numeric
incorrect.
Frequent patient activity.
RESP
waveform
disabled.
RESP waveform amplitude
faint.
ECG module defective.
HR
Measurement
defective.
Possible Cause
SpO
transducer or module
2
defective.
Patient activity.
Extremely strong ambient
light.
The patient is injected with
dye.
Possible Cause
The cuff is folded or air
leak occurs.
Loose
cuff
activity.
Cuff size unsuitable for
patient.
NIBP module defective.
Possible Cause
value
Bad connection of TEMP
transducer.
Quiet the patient.
is
Enable the RESP waveform in
system menu.
Adjust
multiple in RESP menu.
Replace ECG module.
waveforms
Adjust the connection.
Solution
Replace the SpO
Quiet the patient.
Weaken ambient light.
Eliminate dye before measurement.
Solution
Repair the cuff.
or
patient
Tie up the cuff or quiet the patient.
Use appropriate cuff.
Replace NIBP module.
Solution
Connect
transducer well.
- 35 -
Troubleshooting
waveform
amplitude
transducer.
2
and
fix
the
TEMP