Patient Monitor User Manual
4
To zero the sensor:
- Expose the sensor to room air and keep it away from all sources of CO
ventilator, the patient's breath and your own.
- Start up CO
MEASURE
- In the CO
SETUP menu, select ZERO CAL.
2
- The messages indicate: "zero started" , "zero successful". After the zero calibration is
finished, the user can start CO
5 Install the airway adapter at the proximal end of the circuit between the elbow and the
ventilator Y-section.
Figure 17-6 Connecting sensor
SETUP menu, and change WORK MODE from STANDBY to
2
Monitoring.
2
Figure 17-7 Connecting airway adapter
- 138 -
CO
Measuring (Optional)
2
including the
2